跳至主要内容

Difference Between Oxygen Index and Limit Oxygen Index



First, the oxygen index and the limit oxygen index, respectively, what is the meaning, what is the point?The limit oxygen index refers to the minimum oxygen concentration, also known as the oxygen limit and the oxygen index, required for the material to maintain the state of combustion in the oxygen-nitrogen mixture under specified test conditions. It is worth noting that the oxygen index does not refer to the oxygen fraction of oxygen gas mixed with the volume fraction of oxygen, this is the oxygen concentration value.Oxygen index is high that the material is not easy to burn, the oxygen index is low that the material is easy to burn, it is generally believed that the oxygen index <22 is a flammable material, oxygen index between 22 --- 27 is a combustible material, oxygen index> 27 is a flame retardant material.

Second, how to calculate the limit oxygen indexThe limiting oxygen index LO I is expressed as a percentage by volume and is calculated as follows:LO I = cF ten k dWhere: LO I - limit oxygen index,%;CF a test of the last oxygen concentration, take a decimal,%;D a test when the difference between the two oxygen concentration, take a decimal,%;K a coefficient, look-up table.When LO I is reported, the LO I should be calculated to the decimal two bits when calculating the standard deviation e.

oxygen index tester
Third, the limit oxygen index test K value is determinedIf the last five oxygen index values ​​are tested, the first reaction symbol is "X", find the corresponding five measured response symbols in the first column of the table below, and find out from the item (a) The number of K numbers corresponding to the number of "O".
Four, oxygen index tester or limit oxygen index analyzer is used to test whatUsed to test the ultimate oxygen index of the material to evaluate the combustion properties of the material. Suitable materials include homogeneous solid materials, laminates, foams, films and films.
Five, oxygen index tester for what is the standardISO 4589-2, ASTM D2863, GB / T 2406, GB / T 5454

oxygen index tester
Six, oxygen index tester principleThe sample was vertically fixed in the transparent combustion tube of oxygen and nitrogen mixed gas, igniting the top of the sample, observing the burning characteristics of the sample, comparing the continuous burning time of the sample or the burning length of the sample to a given limit , The lowest oxygen depth values ​​expressed as the percentage of oxygen at the time of combustion were measured by a series of tests at different oxygen concentrations.

Read more on http://www.qinsun-lab.com/

评论

此博客中的热门博文

Tinuvin 1130: An Excellent UV Absorber

 In the field of the chemical industry, UV absorbers play a crucial role, particularly in protecting materials from UV damage. Today, we will delve into a highly regarded UV absorber—Tinuvin 1130. Tinuvin 1130 : chemical name: 3-(3-(2H-benzotriazole-2-yl)-4-hydroxy-5-tert-butyl phenyl)propionic acid polyethylene glycol 300 ester; CAS login number: 104810-48-2. This product is an exceptional UV absorber that offers numerous benefits due to its unique properties. One of the standout features of Tinuvin 1130 is its excellent weather resistance. It has a strong ability to absorb ultraviolet rays, particularly those with wavelengths ranging from 290-400nm. This effectively prevents coatings, plastics, rubber, and other materials from being damaged by UV rays, thereby maintaining the performance and appearance stability of these materials and extending their service life. In addition, Tinuvin 1130 exhibits good compatibility. It is easy to emulsify in water-based systems and can be...

Ink Abrasion Tester For Ink Layer Abrasion Test

        Abrasion resistance of ink layer is one important parameter that is used to evaluate the printing quality. If the abrasion resistance of the ink layer is poor, the printed words on the paper or paperboard may become illegible and the part with printed area may be abraded in transportation, storage or in use.                   As a result, the printed words will be not able to be read by the customers, which cause inconvenience or misunderstanding and thus the product image would be compromised. Therefore, it is necessary for the manufacturers to test the abrasion resistance of the ink layer on the paper or paperboard before use.         Often during the shipping or handling process, printed products will be exposed to a degree of abrasion to its surface. Regardless of quality or perfection of the printed product, a scuffed, rubbed off or scratched surface would render it a print defect...

Medical Mask Synthetic Blood Penetration Tester

Surgical Masks Synthetic Blood Penetration Resistance Tester Medical Face Mask Synthetic Blood Penetration Resistance Tester is mainly used in the blood penetration test of masks, which ensures that medical masks can have enough protection to prevent the penetration of blood after splashing from bringing danger to users. It is an essential part of mask test, especially for medical masks. Face Mask Synthetic Blood Penetration Resistance Tester can be used not only to detect synthetic blood penetration resistance of medical masks, but also to detect other textile coating materials. When in use, it can be used to detect different directions only by adjusting the appropriate parameters, thus improving the practicability and working efficiency of the equipment to the greatest extent. Mask Synthetic Blood Penetration Tester  Standards: GB 19083-2010 Technical requirements of medical face masks - Anti-penetration property of 5.5 synthetic blood. YY/T 0691-2008 Clothing for protection agai...